Protea - Protir 204M/C Transportable FTIR
MCERTS APPROVED GAS EMISSIONS MONITORING
Stack Emissions Compliance Testing | Back-up CEM | Suitable for WID and LCPD process for testing under EN 14181 or as a back-up to fixed CEM |
 |
- The Protea 204M/C is a high resolution multi-component FTIR analyser, designed for emissions monitoring of combustion plants, such as those that fall under WID and LCPD.
- The analyser is fitted with an integrated sampling control system that ensures conditioned sample gas passes to the analyser. This system contains temperature and pressure instruments that give a continuous status on the analyser's operation.
- The 204M/C contains an on-board zirconia oxygen sensor. This allows for the collection of oxygen results parallel to combustion gas results and gives the system the unique ability to actively correct for oxygen content online.
- The Protea 204M/C analyser is designed to be run with Protea Analyser Software (PAS). This allows for the continuous collection, labelling and storing of spectral data and concentrations from the FTIR.
- For measurement of combustion gases, PAS runs pre-defined analysis routines for 12 common combustion gases. Each analyser can be calibrated and configured at commissioning and after all maintenance work. Additional components can be added very simply.
- The 204M analyser is unique in that it is designed for both the MCERTS-professional monitoring user and also the plant operator who requires an easy to install and operate back-up to plant fixed CEM units.
|
 |
| |
|
Expert Support and Development:
- Extensive training available
- Full spectroscopic, chemometric and data analysis support by Protea's own UK based experts
- Expert monitoring applications support
- Emergency + QAL2 monitoring backup
- Equipment maintenance packages
- Calibration and support services available
|
 |
| Approved Combustion Gases |
|
Possible Additional Component |
| Component |
Range |
|
Formaldehyde |
| H20 |
0 - 40% |
|
HBr |
| CO2 |
0 - 20% |
|
HCN |
| NOx |
0 - 200 mgm-3 |
|
Benzene |
| SO2 |
0 - 75 mgm-3 |
|
Speciated VOCs |
| HCl |
0 - 15 mgm-3 |
|
|
| NH3 |
0 - 15 mgm-3 |
|
|
| HF |
0 - 50 mgm-3 |
|
|
| CH4 |
0 - 50 mgm-3 |
|
|
Specifications:
| Dimensions and Weight: |
117 x 66.5 x 43cm. 71.3kg |
| Main Power: |
115V, 10A or 230V, 5A, 50/60 Hz (1200 VA startup, 150 VA typical operation) |
| Sample Temperature: |
Gas Cell @ 180°C. Warm-up time 1 hour. Heated sample line and sample probe required |
| Ambient operating temperature: |
+5°C to +40°C |
| Outputs: |
Analogue, 16 channels. Digital, 5 channels |
| Inputs: |
Analogue and digital, depending on analyser specification |
| Resolution: |
Variable resolution FTIR. Maximum resolution 1cm-1 |
| Gas Cell: |
Aluminum gas cell of volume 2.7 litres.
Ambient pressure during normal operation.
Recommended pressure range, -0.8 to 1.2 bar |
| Optics: |
Zinc Selenide beam splitter. Does not require interferometer purge. 6.4m Path length gas cell with pre-aligned coated mirror carriage Barium Fluoride optical gas cell windows (changeable dependant on application –calibration required) |
| Zero Gas: |
99.999% Nitrogen |
Electronics within the analyser allow for the 204M/C to integrate with current industry standard plant systems. The unit can output results of gas concentrations and instrument status levels to plant data control systems, allowing for seamless back-up to fixed CEM systems.
|
|